18
Measured value checkout time
IMPORTANT
This function may not be activated for applications in which it can be assumed that the
measured value will not change.
This function tests whether the measured value varies from that of the sensor (at the measured value input)
within the “Measured value checkout time”. It is assumed that it will do so for an intact sensor.
If the measuring value does not change during this checkout time, the DULCOMETER
®
D1C sets the
control variable to “0” and the alarm relay drops out. The LCD display shows e.g. the message “Check Cl
probe”.
Complete Operating Menu / Description
Correction value pH
Possible values
Initial value Increment Lower value Upper value Remarks
Correcting value off off A pH change < pH 0.1
manual must remain at “manual”
automatic
Type of sensor CLE CLE
CGE/CTE
Standard signal input 4 mA 4 mA
Lower signal limit 0 mA
Scheduled measuring pH 15.45 ... pH 0.01 pH -2 pH 16
range pH -1.45
D1C2-Cl-021-GB
Access to all setting menus can be blocked with an access code !
Only with correction value pH
pH
correcting value
setting ?
manual
4.50 pH
correcting value
range adjustment
4…20 mA
correcting value
range adjustment
20 mA =
4mA =
correcting value
-1.45 pH
15,45 pH
sensortype
CLE
correcting value
Possible values
Initial value Increment Lower value Upper value Remarks
Standard signal input 4 mA 0 mA
lower signal limit 4 mA
Allocated measured
value
lower 0 ppm
upper 2 ppm 0.01 ppm 0.00 ppm 100.00 ppm
Checkout time off 1 s 1 s 9999 s Constant measurement signal
results in message and alarm.
Function off = 0 s
BA_DM_173_03_09_GB.p65 30.03.2009, 12:10 Uhr18