13
Restricted Operating Menu / Description
General Settings
D1C2-Cl-016-GB
Access to all setting menus can be blocked with an access code !
general setting
information
ident-code: D1CA
software version
DxCxxxxxxxxxx
D1C-A1 FW-5.00
alarm relay access c.:
=
=
operating menu:
active
5000
english
reduced
Possible values
Initial value Increment Lower value Upper value Remarks
Alarm relay active active
not active
Access code 5000 1 1 9999
Language as per identity as per identity
code code
Operating menu restricted restricted
complete
Access Code
Access to the setting menu can be prevented by setting up an access code. The DULCOMETER
®
D1C
controller is supplied with the access code 5000 which permits free access to the setting menu. The
calibration menu remains freely accessible even if access to the setting menu is blocked by the code.
BA_DM_173_03_09_GB.p65 30.03.2009, 12:10 Uhr13