PT200MI instruction manual 13
5.2.4 DECIMAL POINT
Transmit Response
[num]<CR><LF> D.P [num]<CR><LF>
Options for [num]
0 = No DP 1 = XXXX.X 2 = XXX.XX 3 = XX.XXX 3 = X.XXXX
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
MULT [num]<CF><LF>
YES
5.2.5 MULTIPLIER (It is only possible to modify the multiplier when no decimal point is set)
Transmit Response
[num]<CR><LF> MULT [num]<CR><LF>
Options for [num]
1 = 1 10 = 10
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
d [num]<CF><LF>
YES
5.2.6 DIVISION
Transmit Response
[num]<CR><LF> d [num]<CR><LF>
Options for [num]
1 = 1 2 = 2 5 = 5
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
MAX.CAP [num]<CF><LF>
YES
5.2.7 MAXIMUM CAPACITY
Transmit Response
[num]<CR><LF> MAX.CAP [num]<CR><LF>
Options for [num]
500 1000 1200 1500 2000 2500 3000 4000
5000 6000 8000 10000 12000 15000 20000 25000
30000 40000 50000 60000 80000 100000
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
BAUD [num]<CF><LF>
YES
Note that this is the maximum displayed reading without the decimal point showing.
If there is any error in the maximum capacity, step width and multiplier, the error message “ERROR 1”
will be sent out, PT200MI will return to 5.2.4.
5.2.8 RS232/RS485 BAUD RATE
Transmit Response
[num]<CR><LF> BAUD [num]<CR><LF>
Options for [num]
2400 4800 9600 19200
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
UNIT [string]<CF><LF>
YES