PT200MI instruction manual 14
5.2.9 UNIT
Transmit Response
[string]<CR><LF> UNIT [string]<CR><LF>
Options for [string]
<SP>t = ton kg = kg
(Note <SP> is a single space required before the t)
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
Z.RANGE [num]<CF><LF>
YES
5.2.10 ZERO RETURN RANGE
Transmit Response
[num]<CR><LF> Z.RANGE [num]<CR><LF>
Options for [num]
1 2 3 4 5 6 7 8 9 10
(NOTE these numbers are actual % of range)
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
D.FILTER [NUM]<CF><LF>
YES
5.2.11 DIGITAL FILTER
Transmit Response
[num]<CR><LF> D.FILTER [num]<CR><LF>
Options for [num]
0 2 4 8 16 32 64 128 256 512
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
DSP RATE [NUM]<CF><LF>
YES
5.2.12 DISPLAY UPDATE RATE
Transmit Response
[num]<CR><LF> DSP RATE [num]<CR><LF>
Options for [num]
1 = 1 times/sec 4 = 4 times/sec 8 = 8 times/sec 16 = 16 times/sec
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
BCD RATE [NUM]<CF><LF>
YES
5.2.13 BCD OUTPUT RATE
Transmit Response
[num]<CR><LF> BCD RATE [num]<CR><LF>
Options for [num]
4 = 4 times/sec 8 = 8 times/sec 16 = 16 times/sec 20 = 20 times/sec
60 = 60 times/sec 80 = 80 times/sec 100 = 100 times/sec 200 = 200 times/sec
N<CR><LF> to move to the next setting.
R<CR><LF> to return to normal weighing.
ID. NO. [num]<CF><LF>
YES