OPER
ATING MANUAL FOR SELECT ANALYST
Issue: A, part number R041118
11
4 About your water purifier
The Purite Select Analyst purifies potable water. The water is purified by a
combination of processes including; ion-exchange, activated carbon and
reverse osmosis.
The purified water produced by the unit complies with latest standards for
laboratory grade purified water.
4.1 The Select Analyst range of units
GP: Generation unit (with no close coupled tank).
IT: Generation unit with close coupled 20 litre water storage
tank
GP BP: Generation unit with feed water boost pump
BP/IT: Generation unit with close coupled water storage tank and
feed water boost pump
Each of the above variants is available with 4 different output flow rates:
SA 40 6 litres per hour
SA 80 12 litres per hour
SA160 24 litres per hour
SA 320 48 litres per hour
Please note that the above outputs are based on an 8 hour working day at
60psi (4 bar) inlet pressure, 25
o
C and maximum of 400 ppm total dissolved
solids.
The unit type is detailed on its serial plate which can be found on the
reverse side of the front cover.
The Select Analyst is provided with an installation kit for quick installation.
Both media packs are provided and packaged separately. Refer to Section
9.1 and 9.6.
4.2 Key components of the Select Analyst
4.2.1 Touch screen display
Operation of the Select Analyst is controlled solely through the touch
screen display. System schematics are displayed for easy
understanding of the unit’s process.
4.2.2 Close coupled tank
The close coupled tank holds 20 litres of purified water. To dispense