OPER
ATING MANUAL FOR SELECT ANALYST
Issue: A, part number R041118
22
o The filter is located in the exposed panel on the right hand side
of the unit.
To obtain access to the filter screen unscrew the black knurled top anti-
clockwise. The screen will then be exposed. Pull the screen away
from the housing and clean away any debris.
Once cleaned, replace the screen and black top. Replace the side
panel and cover using the reverse procedure and switch on the water
supply. Start the unit from the power on screen
6.3 Sanitisation procedure.
Sanitisation of the unit should be carried out at least once every 6 months.
Switch the unit off by pressing the STOP button.
The unit will now have reverted to the POWER ON screen.
Turn off the water supply to the unit.
Open the front door cover to reveal the media cartridge packs.
Remove both media cartridge packs.
To sanitise the unit insert a cleaning pack adaptor (Purite part number
L998549) into each of the media pack locations.
Turn on the water supply and start the unit.
Allow the unit to fill the tank. When the tank is full the unit will stop.
Allow the unit to stand with a full tank for two hours. Drain the tank
using the bib tap located on the front of the tank.
The unit will automatically restart and begin to fill the tank. When the
tank is full drain it and allow it to fill once more.
When full switch the unit off, drain the tank and fit new media packs.
Restart the unit and allow it to fill the tank. When filled, switch the unit
off and drain the tank again.
Once the tank is drained check for residual sanitising chemicals.
Only when you are happy that all chemicals have been flushed from
the system should the unit be used. Test kits are available, contact
Purite for details.