8.3 Feedwater supply requirements
Free chlorine 0.5 - 1.0 ppm as Cl
2
Total dissolved solids (max) 1000 ppm
Temperature 1 - 40
o
C (33.8 – 103.9
o
F)
Feedwater pressure (max) (90 psi)
Feedwater pressure (min) (30 psi)
Design pressure (60 psi)
Contact your supplier if the feedwater supply is outside these limits.
8.4 Electrical Supply
Your Select HP will require a single phase electrical supply that meets the
requirements below:
Electrical
supply
KW
rating
Current
draw (amps)
External fuse
rating
Single phase
100-240v
+
10% 50 Hz
and Earth
0.1
1.0
Amps
(conforming to
BS1362)
Ensure that the mains supply is provided via a convenient switched outlet and
is protected with a 5 amp fuse.
8.5 Unit positioning
The Select HP has been designed to be mounted on a bench or wall.
Bench Mounted
Locate the unit on a suitable work surface, ensuring access to tank
dispense bib tap and sufficient clearance to rear of the machine for all
water and electrical connections.
All water feed connections are found on the rear of the unit. Using the
installation kit provided attach the water feed, drain and overflow tubing.