Testing InfoFrames (HDMI only)
The HDMI analyzer can analyze InfoFrame data from an incoming HDMI signal. With this
capability, an HDMI transmit device can be tested for its ability to transmit InfoFrame
packets correctly.
Testing HDMI transmit device InfoFrame capability
The 802 generator can monitor the InfoFrame content of the received HDMI signal.
To view current HDMI InfoFrame contents received from an HDMI transmitter:
1. Connect an HDMIdisplay to the HDMI Tx connector on the generator.
2. Connect the HDMI source to the HDMI Rx connector on the generator.
3. Setup the generator for HDMI output (see page 147).
4. If desired, monitor the video signal received on the HDMI Rx connector (see
Monitoring HDMI analyzer signal input” on page 110).
5. Using the Format knob, select an appropriate format for the connected HDMI/DVI
display.
Note: To automatically configure the Format knob list based on the EDID structure of
a DDC-compliant display, hold down the R key, then press and release the ACS key.