21.2. Responses format for commands sent from computer level
On receipt of a command, the terminal sends a response in one of the following
formats:
XX_A CR LF command understood and in progress
XX_D CR LF command carried out (appears only after the command XX_A)
XX_I CR LF Command understood but not accessible at this moment
XX _ ^ CR LF Command understood but max range is exceeded
XX _ v CR LF Command understood but min range is exceeded
XX _ OK CR LF Command carried out
ES_CR LF Command not recognised
XX _ E CR LF
An error occurred on command carrying out (time limit
exceeded while waiting for stable measurement result), time
limit is balance’s characteristic parameter
XX - stands for a name of sent command
_ - stands for spaces
21.3. Commands description
Zero the balance
Format: Z CR LF
Accessible responses:
Z_A CR LF
Z_D CR LF
- command understood and in progress
- command carried out
Z_A CR LF
Z_^ CR LF
- command understood and in progress
-
command understood but zeroing range exceeded
Z_A CR LF
Z_E CR LF
- command understood and in progress
-
time limit exceeded while waiting for stable measurement result
Z_I CR LF
- command understood but not accessible at this moment