Filter                                 Average                           
Value release         Fast &Reliable  
----------------------------------------  
Deviation for Max.             -0.118 g                               
Repeatability                    0.0088 g                          
Signature                                                               
  
........................................ 
 
Procedure: 
To initiate parameter <P6.11 GLP AUTOTEST > press 
 button. The balance software launches 
the GLP autotest procedure and it is carried out automatically from now on to the end.  
You can stop the procedure at any time by pressing 
 button of the GLP process window.  
Once the procedure is finished, the value of standard deviation of all measurements is shown in the 
main display, and message <RESULT> in the bottom line, the final report is printed out automatically 
(as shown above). It is possible to print out the report again by pressing 
 button. 
Report is not saved, therefore when abandoning this level, it is deleted from the balance volatile 
memory.  
To abandon the window press 
 button and the program returns to the main menu. To return to 
weighing, press 
 button repeatedly. 
 
7.  ADJUSTMENT 
In order to ensure the highest weighing accuracy, it is recommended to periodically introduce a 
corrective factor of indications to balance memory, the said factor must be referred to a mass 
standard. In other words, balance adjustment shall be performed from time to time.  
 
Adjustment should be carried out: 
•  Before the beginning of weighing procedure, 
•  If long breaks between following measuring series occur, 
•  If temperature inside the balance changes more than: 1°C or 2°C for AS series balances or 2°C 
for PS series balances. 
 
Types of adjustment: 
•  Internal automatic adjustment 
•  Manual internal adjustment  
•  Adjustment with an external weight of declared mass which cannot be modified or of any mass, 
but not lower than 30% of maximum range.  
 
CAUTION! 
In the case of verified balances (with an internal automatic adjustment system) only automatic internal 
adjustment and manual internal adjustment are available. Remember to carry out the adjustment 
process when there is no load on the pan! When the weighing pan is loaded, command <RANGE 
EXCEEDED> is displayed. In such a case remove the load and restart the adjustment process. Adjustment 
process can be aborted if necessary by pressing Esc button at any time during the process.  
 
7.1. INTERNAL ADJUSTMENT 
Adjustment process can be initiated automatically and manually.