- 56 - 
 
After weighing process completion, the measurement result is locked and automatically printed.  
In order to finish the measurement, press 
 button. 
The program automatically returns to the main window of Animal weighing mode.  
11.7.3.  Running the Process Automatically – Means of Operation 
To start the process of weighing automatically, enter the mode settings, select an <AUTOSTART> 
parameter and set <YES> value, follow description form the previous paragraph. Additionally set 
parameters <AVERAGING TIME> and <THRESHOLD>. 
In order to carry out the measurement in a container (TARE) for this mode, select the option <ENTER 
TARE>, for the option overview to be found in weighing mode options description. 
To start the process (after setting the options), enter the weight of the container, then put the 
container on the pan and place the object that is being weighed in the container. The balance 
automatically begins the process of measurement after exceeding the set mass threshold.  
 
After weighing process completion, the measurement result is locked and automatically printed.  
In order to finish the measurement, press 
 button. 
The program automatically returns to the main window of Animal Weighing mode.  
 
11.8. DENSITY OF SOLIDS 
Solids  Density  is a working mode enabling determination of density of a representative sample 
material.  
In order to operate this mode, an optional density kit (supplementary equipment), appropriate to 
a model of used balance, is required. Prior to installing the kit it is necessary to remove the weighing 
pan and a draft shield. In the place of the weighing pan assemble the weighing pan stand (1) and 
place beaker’s basis (10) on it.  
 
 
Kit intended for PS balances featuring 128x128 mm weighing pan: 
2 
Bottom weighing pan for solids 
density determination  
8 
Thermometer holder