- 68 - 
 
. 
. 
. 
5. Net 
 Tare 
 
 
 
 5.000g 
 0.000g 
- - - - - - - - - - - - - - -  
- - - - - - - - - - - - - - - 
In order to continue adding process, press   button on printing out the sum. Home screen of the 
main mode is displayed. 
If samples are measured with packagings, load the weighing pan with the packaging, wait for 
stabilization and press 
 button. 
 
CAUTION:  for ADDING mode, on pressing 
 button (confirming mass of measured sample) data 
connected with the measurement is printed. The following data can be printed: Measurement 
number,  NET mass  (in calibration unit), TARE mass  (in current unit), GROSS mass  (in current 
unit), and current result  (in  current unit). Data to be printed is to be set in GLP PRINTOUT 
parameter  -  see point 8 of this user manual. Other information (except the above-listed), such as: 
USER, PRODUCT, DATE, TIME, VARIABLE 1, VARIABLE 2, CALIBRATION REPORT, NON-
STANDARD PRINTOUT are not printed. 
11.14.  PIPETTES CALIBRATION 
Caution! Function valid for AS R series balances exclusively. 
 
Pipettes calibration function applies to fixed volume pipettes and adjustable volume pipettes. During 
volume testing procedure, the software determines accuracy and repeatability errors. In the case of 
pipettes with adjustable volume, errors for Max, Min and ½ Max volume are estimated. 
All pipettes are tested for adherence to  requirements of PN-EN ISO 8655:2003. During the tests 
dosing repeatability and accuracy are monitored. 
 
In order to ensure the highest accuracy of pipettes calibration process, maintain the following 
ambient conditions at a workstation:  
•  Ambient temperature of a pipette, tips and liquid should be kept between 20°C -  25°C with 
change rate during testing within ± 0.5°C, 
•  Relative humidity 50 - 75%; 
and  
•  Use distilled water for pipettes calibration processes, 
•  Make sure that a pipette, tips and distilled water are thermally stabilized in the room intended 
for weighing operation performance. The  reference standard advises that minimum 
acclimatization time for above mentioned is 2 hours.  
 
Prior to pipettes calibration start, it is necessary to install a dedicated set inside the draft shield. The 
set is not a standard balance equipment. The picture below presents set installation. Evaporation ring 
minimizes measurement errors being a result of evaporation of liquid occurring in course of the 
weighing process.