Command reference
R&S
®
ZNA
1031User Manual 1178.6462.02 ─ 20
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example:  See CALCulate<Chn>:RIPPle:FAIL?
Manual operation:  See "Ripple Check" on page 440
7.3.1.19  CALCulate:SMOothing...
The CALCulate:SMOothing... commands provide the settings for trace smoothing.
CALCulate<Chn>:SMOothing[:STATe]............................................................................1031
CALCulate<Chn>:SMOothing:APERture........................................................................ 1031
CALCulate<Chn>:SMOothing[:STATe] <Boolean>
Enables or disables smoothing for trace no. <Chn>.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<Boolean> ON | OFF - smoothing on or off.
*RST:  OFF
Example:  See CALCulate<Chn>:SMOothing:APERture
Manual operation:  See "Smoothing" on page 417
CALCulate<Chn>:SMOothing:APERture <SmoothAperture>
Defines how many measurement points are averaged to smooth the trace.
Suffix: 
<Chn>
.
Channel number used to identify the active trace
Parameters:
<SmoothAperture> Smoothing aperture. A smoothing aperture of n % means that
the smoothing interval for each sweep point i with stimulus value
x
i
 is equal to [x
i
 - span*n/200, x
i
 + span*n/200], and that the
result of i is replaced by the arithmetic mean value of all mea-
surement points in this interval.
Range:  0.05% to 100%.
*RST:  1
Default unit: %
SCPI command reference