EasyManua.ls Logo

R&S ZNA50 - Page 1172

R&S ZNA50
1709 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Command reference
R&S
®
ZNA
1172User Manual 1178.6462.02 ─ 20
Manual operation: See "Restart Sweep" on page 512
INITiate<Ch>[:IMMediate]:SCOPe <Scope>
Selects the scope of the single sweep sequence. The setting is applied in single sweep
mode only (INITiate<Ch>:CONTinuous OFF).
This command is required in compatibility modes only (see SYSTem:LANGuage).
Suffix:
<Ch>
.
Channel number
Parameters:
<Scope> ALL | SINGle
INITiate<Ch>[:IMMediate][:DUMMy] starts a single
sweep in all channels or in the referenced channel <Ch> only.
Example: See INITiate<Ch>:CONTinuous
Manual operation: See "Sweep All Channels/Sweeps" on page 514
INITiate<Ch>:CONTinuous <Boolean>
Qualifies whether the analyzer measures in single sweep or in continuous sweep
mode.
Suffix:
<Ch>
.
Channel number.
This suffix is ignored in the "ZVR" and "ZVABT" compatibility
modes (SYSTem:LANGuage 'ZVR' | 'ZVABT').
Parameters:
<Boolean> ON - the analyzer measures continuously, repeating the current
sweep.
OFF - the measurement is stopped after the number of sweeps
defined via [SENSe<Ch>:]SWEep:COUNt. INITiate<Ch>[:
IMMediate][:DUMMy] initiates a new measurement cycle.
*RST: ON
SCPI command reference

Table of Contents

Other manuals for R&S ZNA50

Related product manuals