1517User Manual 1178.6462.02 ─ 20
SYSTem:TRESet[:STATe] <Boolean>
Defines the behavior of the INITiate<Ch>[:IMMediate][:DUMMy] and
INITiate[:IMMediate]:ALL.
Parameters:
<Boolean> OFF (0)
Previous trace data are preserved.
ON (1)
Previous trace data are deleted.
This is a global setting that is not affected by an instrument
reset. Factory default is OFF (0).
Manual operation: See "Restart: Set all Traces to 0" on page 792
SYSTem:TSLock
<Type>
Locks the touchscreen functionality of the R&S ZNA, e.g. in order to prevent inadver-
tent entries during remote control.
Parameters:
<Type> OFF | DIAGrams | SCReen
OFF touchscreen (e.g. the remote screen) remains active. You
can use the buttons in "Remote" softtool panels. If you switch
back to manual control, the drag and drop functions (e.g. for
markers) are still available.
DIAGrams lock the drag and drop functions in diagrams .
SCReen lock the entire screen including the remote screen
buttons.
*RST: n/a (*RST does not affect the tochscreen lock set-
ting).
Example:
SYSTem:TSLock SCReen
Lock the entire screen.
Manual operation: See "Enabled / Lock Diagrams / Lock Screen" on page 784
SYSTem:TTLout<Pt>:STATus[:STATe] <Boolean>
Defines the default values of CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe] for
new traces.
Suffix:
<Pt>
.
1 - TTL out pass 1 (pin 13 of User Port connector)
2 - TTL out pass 2 (pin 14 of User Port connector)
Parameters:
<Boolean> OFF (0)
New traces are created with
CALCulate<Chn>:LIMit:TTLout<Pt>[:STATe] set to OFF
(0).