Command reference
R&S
®
ZNA
984User Manual 1178.6462.02 ─ 20
Example: See CALCulate:MARKer:COUPled[:STATe] on page 983
Manual operation: See "Coupling Type" on page 472
CALCulate:MARKer:FUNCtion:BWIDth:GMCenter <arg0>
Defines how bandfilter searches calculate the center frequency of the passband or
stopband.
Parameters:
<arg0> ON – use geometric mean of lower and upper band edge
OFF – use arithmetic mean
*RST: n/a (*RST does not affect the calculation rule; the
factory setting is ON/geometric mean)
Manual operation: See "Geometric Calculation of Bandfilter Center" on page 796
CALCulate<Chn>:MARKer[:STATe]:AREA <HorizontalPos>, <VerticalPos>
Moves the marker info field for the active trace <Chn> to one of nine predefined posi-
tions in the active diagram.
Suffix:
<Chn>
.
Channel number used to identify the active trace
Parameters:
<HorizontalPos> LEFT | MID | RIGHt
Horizontal position
<VerticalPos> TOP | MID | BOTTom
Vertical position
Example: See CALCulate<Chn>:MARKer<Mk>:Y
Manual operation: See "Mkr 1 ... Mkr 10" on page 452
CALCulate<Chn>:MARKer:DEFault:FORMat <OutFormat>
Defines the default marker format of the selected channel's active trace.
New markers will be formatted with the default marker format; previously existing mark-
ers will be reformatted if (and only if) their marker format is set to (Trace) DEFault
(using CALCulate<Chn>:MARKer<Mk>:FORMat).
Suffix:
<Chn>
.
Channel number used to identify the active trace
SCPI command reference