CALCulate<Chn>:MARKer<Mk>:THReshold <Value>
CALCulate<Chn>:MARKer<Mk>:REFerence:THReshold <TargetSearchVal>
Defines a marker-specific threshold value for (single) peak searches with the related
marker (see "Marker addressing" on page 681) .
Use CALCulate<Chn>:MARKer<Mk>:THReshold:STATe or CALCulate<Chn>:
MARKer<Mk>:REFerence:THReshold:STATe to activate it.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number (ignored for reference markers)
Parameters:
<Value> Threshold value
The unit is derived from the active trace format and cannot be
modified. Changing the trace format resets the threshold to a
format-specific default value.
Manual operation: See "Threshold Settings" on page 463
CALCulate<Chn>:MARKer<Mk>:THReshold:STATe <Active>
CALCulate<Chn>:MARKer<Mk>:REFerence:THReshold:STATe <Boolean>
Activates or deactivates the marker-specific threshold for (single) peak searches with
the related marker (see "Marker addressing" on page 681) .
Use CALCulate<Chn>:MARKer<Mk>:THReshold or CALCulate<Chn>:
MARKer<Mk>:REFerence:THReshold to set the threshold value.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number (ignored for reference markers)
Parameters:
<Active> Boolean
Manual operation: See "Threshold Settings" on page 463
CALCulate<Chn>:MARKer<Mk>:TYPE <Mode>
CALCulate<Chn>:MARKer<Mk>:REFerence:TYPE <Mode>
Sets the marker mode for the related marker (see "Marker addressing" on page 681).
The marker must be created before using CALCulate<Chn>:MARKer<Mk>[:STATe]
ON or CALCulate<Chn>:MARKer<Mk>:REFerence[:STATe] ON.
Suffix:
<Chn>
.
Channel number used to identify the active trace
<Mk> Marker number (ignored for reference markers)
Parameters:
<Mode> NORMal | FIXed | ARBitrary