EasyManua.ls Logo

RayTek MI Miniature Infrared Sensor - Page 86

RayTek MI Miniature Infrared Sensor
95 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
ProgrammingGuide
78 MI
Description Char Format P B S Legal values Factory
default
LCD
Source: emissivity /
setpoint for alarm output
ES X * * I=constant
number
(E=0.950)
E=external
analogous input
FTC1
D= E/XS digital
selected FTC1-3
I
Presel. emissivity value EV n.nnn * * 0.100 - 1.100
Valley hold time(4) F nnn.n * * * 0.000 - 998.9 s
(999 = infinite)
000.0 s V
Average time G nnn.n * * * 000.0 – 999 s 000.0 s A
Top of mA/mV range H nnn.n * * * (1) (7) H
Sensor / head ambient I nnn.n * * in current scale
(°C/°F)
Switch panel lock J X * * L=locked
U=unlocked
unlocked
Alarm output control K N * * 0=off
1=on
2=Target.; norm.
open
3=norm. closed
4=Head; normal
open
5=norm. closed
7=sensor / head
ambient
?
Bottom of mA/mV range L nnn.n * * * (1) (8) L
Output voltage O n.nnn * * 0.000 – 5.000
voltage in V
6=controlled by
unit
6
Output current O nn.nn * * 0.00 – 20.00
current in mA
21=over range
60=controlled by
unit
60 L
Peak hold time (4) P nnn.n * * * 000.0 998.9 s
(999 = infinite)
000.0 s P
Power/AD value Q nnnn * *

Table of Contents