ProgrammingGuide
MI 79
Description Char Format P B S Legal values Factory
default
LCD
Presel. setpoint / relay
function
SV nnn.n (1)
Target temperature T nnn.n * * in current scale
(°C / °F)
Temperature unit U X * * * C / F C U
Poll / Burst mode V X * * P = poll
B = burst
Poll mode
Burst string contents X$ *
Multidrop address XA nnn * * 000 – 032
000 = single unit
unchanged
(preset: 0)
Bottom temperature of
range
XB nnn.n *
Restore factory defaults XF *
Transmission XG n.nnn * * * 0.100 - 1.000 1.000 T
High temperature of
range
XH nnn.n *
Sensor initialization XI n * * * 1 after Reset,
0 if XI = 0
Box temperature XJ nnn.n * * in current
scale(°C / °F)
FTC 3 trigger / hold XN X * * T = trigger
H = hold
T
Analog output mode XO n * * 0 = 0 – 20 mA
4 = 4 – 20 mA
5 = TCJ
6 = TCK
9 = mV
9
Firmware revision XR * e.g. 1.01
Setpoint / relay function
(10)
XS nnn.n * * (1) 250°C
Trigger XT n * * 0 = inactive
1 = active
0
Unit identification XU * e.g. !XUMILT4
Serial number XV * e.g. 98123
Advance hold hysterese
(4)
XY nnn.n * *
Head calibration (9) XZ * *
(1) LT:‐40to600°C(40to1112°F)
(2) n=number,X=uppercaseletter

Table of Contents