EasyManua.ls Logo

Raytheon ST60 - 4.2 User calibration; Starting User calibration

Raytheon ST60
60 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
28
ST60 Tridata Instrument Owner’s Handbook
4.2 User calibration
The User calibration procedures enable you to:
Set the required units for depth readings.
Set the offset for depth readings, i.e. determine whether depth
readings are from the keel of the vessel or from the water line.
Lock the shallow alarm.
Set the required units for speed readings.
Set the speed resolution.
Select the log distance units
Either calibrate the speed reading to Speed Over Ground (SOG) or
manually apply a calibration factor, to obtain correct speed through
the water.
Select temperature units.
Calibrate for correct temperature readings.
Set timer alarm buzzer on or off.
Separate routines are provided for the User calibration of the depth and
speed functions. To carry out either of these routines:
1. Power up the ST60 Tridata instrument.
To start User calibration
hold down
and
for approximately 2 seconds
speed
depth
speed
press either
depth
or
Starting User calibration
D4416-1
see illustration
User calibration - depth
see illustration
User calibration - speed
User calibration
entry screen
040_2c04.p65 09/08/99, 09:3228

Table of Contents

Other manuals for Raytheon ST60

Related product manuals