Real MULTIDEC MD090-1301BTR - Page 24

Real MULTIDEC MD090-1301BTR
33 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
24
8
I.F.
Low pressure
Cyclic (every 4
seconds)
Invalid initial
inflation alarm
Incorrect filling
9
A.F.
Low pressure
Cyclic (every 4
seconds)
Invalid full
inflation mode
alarm
Incorrect
maximum
mattress filling
10
L.P.
Low pressure
Cyclic (every 4
seconds)
Low pressure
alarm
Low pressure
11
H.P.
service
Cyclic (every
4,5 seconds)
High pressure
alarm
High pressure
12
H.t.
service
Cyclic (every
4,5 seconds)
High ambient
temperature
alarm
High
temperature
13
U.I.
service
Cyclic (every
4,5 seconds)
Air valve 1
incorrect
position alarm
Air valve 1
failure
14
L.b.
service
Cyclic (every
15 seconds)
Low battery
alarm
The battery is
charging while
pump operation
15
S.I.
service
Cyclic (repeat)
Service
indicator
16
C.u.
none
none
Calibration not
completed
Calibration not
completed
17
C.C.
none
none
Calibration
Complete
Finish
calibration
18
L. H.
none
none
Headrest uplift
high
Headrest uplift
>45° (±15°)
19
L.L.
none
none
Headrest uplift
low
Headrest uplift
>20° (±15°)