Manufacturer’s declaration-electromagnetic immunity
Test specifications for ENCLOSURE PORT IMMUNITY to RF wireless communications equipment
The device(s) is intended for use in the electromagnetic environment (for home and professional healthcare) specified below.
The customer or the user of the device(s) should assure that it is used in such an environment.
IMMUNITY TEST
LEVEL
(V/m)
Compliance
LEVEL
(V/m)
(for home and
professional
healthcare)
Pulse
modulation b)
18 Hz
FM c)
±5 kHz deviation
1 kHz sine
Pulse
modulation b)
217 Hz
GSM 800/900,
TETRA
800,
iDEN 820,
CDMA 850,
LTE Band 5
Pulse
modulation b)
18 Hz
GSM 1800;
CDMA
1900;
GSM 1900;
DECT;
LTE Band 1, 3,
4, 25;
UMTS
Pulse
modulation b)
217 Hz
Bluetooth,
WLAN,
802.11 b/g/n,
RFID 2450,
LTE Band 7
Pulse
modulation b)
217 Hz
Pulse
modulation b)
217 Hz
NOTE If necessary to achieve the IMMUNITY TEST LEVEL, the distance between the transmitting antenna and the ME
EQUIPMENT or ME SYSTEM may be reduced to 1 m. The 1 m test distance is permitted by IEC 61000-4-3.
a)
For some services, only the uplink frequencies are included.
b)
The carrier shall be modulated using a 50 % duty cycle square wave signal.
c)
As an alternative to FM modulation, 50 % pulse modulation at 18 Hz may be used because while it does not represent
actual modulation, it would be worst case.