Control Tests
When to Perform a Control Test
Performing a control test lets you know
the meter and test strips are working
properly. You should perform a control test
when:
you open a new test strip box.
you left the test strip container open.
you think the test strips are damaged.
you want to check the meter and test
strips.
the test strips were stored in extreme
temperatures, humidity, or both.
you dropped the meter.
your test result does not match how you
feel.
you want to check if you are performing
the test correctly.
About the Control Solutions
Refer to the control solution package
insert for detailed use instructions and
control solution storage conditions.
Use only control solutions compatible
with the ReliOn Platinum system.
Close the control solution bottle tightly
after use.
Write the date you open the control
solution bottle on the bottle label. The
control solution must be discarded 3
months from the date the control
solution bottle was opened (discard
date) or on the Use By date on the bottle
label, whichever comes first.
Do not use control solution that is past
the Use By or discard date.
The meter automatically recognizes the
dierence between the control solution
and blood.
The control solution can stain fabric.
Remove stains by washing with soap
and water.
The control results are not displayed in
memory.
Control solution can be ordered
separately by contacting Customer
Service at 1-800-819-1097.
Performing a Control Test
NOTE
The ranges for Level 1 and Level 2
controls are printed on the test strip
container label.
The meter turns o 90 seconds after
a successful test or 15 seconds after
the test strip is removed, provided no
other action is taken.
You need the meter, a test strip, and
control solution Level 1 or Level 2.
1
Check the Use By
date on the test strip
container. Do not use
test strips past the
Use By date.
2
Remove a test strip
from the test strip
container.
Close the cap tightly.