19
Control Tests
3
8
or
OK appears if the control result is within
range.
Err appears if the control result is out of
range.
9
Wipe the tip of the
bottle with a tissue.
Cap the bottle tightly.
Remove and discard
the used test strip.
Understanding Out‑of‑Range Control Results
WARNING
The control range applies only to a control result. It only indicates that the test strips
and meter are working properly. Do not use a control result to interpret blood
glucose results.
If the control result is out of range, do not use the meter until you solve the problem.
Check the following list to help solve the problem.
Troubleshooting Checks Action
Were the test strips or control
solutions expired?
Discard the test strips or control solution if
either is past the Use By date. If the control
solution was opened more than 3 months ago,
discard it. Repeat the control test with an
unexpired test strip and an unexpired control
solution.
83561_09410155001A EN ReliON.indd 1983561_09410155001A EN ReliON.indd 19 3/23/2022 11:24:07 AM3/23/2022 11:24:07 AM