BiPAP Vision Ventilatory Support System
SERVICE MANUAL

Table of Contents

Questions and Answers

  • H
    hernandezmatthewSep 12, 2025
    What to do if Respironics BiPAP Vision Medical Equipment ventilator operation is interrupted, and the display shows the Exhalation Port Test screen?
    • W
      Whitney HamiltonSep 12, 2025
      If the ventilator operation is interrupted and the display shows the Exhalation Port Test screen, especially for software versions 11.0 - 11.11, 12.0 - 12.7, and 13.0 - 13.7 after a power loss of approximately 10 seconds or longer, exit the screen and restart normal operation. Also, check the power source and power cord connections.
  • C
    cynthiamarquezSep 13, 2025
    What to do if Respironics BiPAP Vision Medical Equipment touch pad does not respond to selection?
    • C
      Christopher PowellSep 13, 2025
      If the touch pad does not respond to selection, it could be due to a defective touch pad or DC board. First, check the connections. If that doesn't work, replace the touch pad. If the issue persists, replace the DC board.
  • C
    Cory CabreraSep 15, 2025
    How to fix Respironics BiPAP Vision when rotary encoder does not adjust selection?
    • R
      Randy GrahamSep 16, 2025
      If the rotary encoder does not adjust the selection, it may be due to a defective rotary encoder or DC board. Try replacing the rotary encoder first. If the problem continues, replace the DC board.

Summary

Warnings, Cautions, and Notes

Warnings

Identifies hazardous conditions that could cause injury or damage.

Cautions

Highlights conditions that could cause device damage or shorten service life.

Theory of Operation

PSS Board

Details the Power Supply Subsystem (PSS) board functions and features.

Periodic Maintenance

Internal Alarm Battery

Information on the internal alarm battery, charging, and error codes.

Component Removal;Installation

MC Board

Steps for removing and reinstalling the Main Control (MC) board.

Troubleshooting

Troubleshooting Common Problems

Describes common problems, their causes, and corrective actions.

Alarm Troubleshooting

Lists alarms, their possible causes, and corrective actions.

Troubleshooting Check Vent Error Codes

Flowchart and table for troubleshooting Check Vent error codes.

Troubleshooting Vent Inop Error Codes

Flowchart and table for troubleshooting Vent Inop error codes.

Testing and Calibration

System Final Test

Verifies ventilator operation and accuracy of measurements.

Appendix B Specifications

Respironics BiPAP Vision Specifications

General IconGeneral
Humidifier CompatibilityIntegrated heated humidifier
Device TypeBiPAP
ModesS, CPAP
Ramp Time0 to 45 minutes
Rise Time1 to 5
AlarmsHigh Pressure, Low Pressure, Apnea, Power Failure, Circuit Disconnect
DisplayLCD
Power Supply100-240V AC, 50/60 Hz
Weight4.5 lbs