57
The values of PULSE DURATIONS, PULSE repetition frequencies and amplitudes, including
any d.c. component, shall not deviate by more than ± 20 % when measured with a load
resistance within the range specified.
If confirmation is required that Revitive works within its Essential Performance aer a
certain period of time, contact the manufacturer
Output Specifications for Transcutaneous Electrical Nerve Stimulation (TENS):
Waveform Biphasic
Shape Rectangular
Maximum Output Voltage: @ 500Ω 50Vp
Pulse Duration: 100µs
Net Charge (per pulse @ 500Ω)
0μC
Maximum Average Power Density (@ 500Ω) 4.19 mW/cm
2
Output Specifications for Electrical Muscle Stimulation (EMS):
Waveform Monophasic
Shape Square symmetrical with polarity reversal
Maximum Output Voltage: @500Ω Foot: 26 - 35Vp Body: 18 - 32.6Vp
Pulse Duration: Foot: 450 or 970µs Body: 450µs
Net Charge @ 500Ω [0.001]mC
Maximum Power Density @ 500Ω Foot: 0.414 mW/cm² Body: 1.52-3.82 mW/cm²
ON Time Foot: 1.90 - 8.30s Body: 1.90 -33.0s
OFF Time Foot: 1.00 - 1.50s Body: 1.00 - 47.0s
Technical Specications