Service Guide for DM3000 Series
3-3
Performance Test
You can select two different methods when testing by using the equipments listed in
table 3-1: Quick test and Common test. The detailed explanations are given below.
Quick Test
Quick test is an eclectic project which can either speed up test or ensure the test
results achieve to high confidence which contains parts of test items within Common
test, additional, “Q” is designated for its short (in Common test).
NOTE
Quick test is not applicable for equipments with some abnormal components. The
equipment failing the quick performance check must be used after calibrating or
repairing.
Common Test
Common test is recommend be done once you got the instrument so as to do a
complete evaluate for its performance. Generally, the test result is valid in 90-days. If
the performance test is performed in 24-hours after calibration, the valid period is
24-hours.
NOTE
It is necessary to do the test again when exceed valid period since first successful
test. The equipment failing test should not be used until have been calibrated or
repaired.

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals