Optional AC Voltage Test
In optional AC voltage test, input the standard signal the test equipment outputs into
the input terminal of tested equipment; and compare the test results with the error
shown in the table 3-12 and 3-13.
Table 3-12 Accessional AC Voltage Test Error (DM306x)
Test Signal
Input Frequency
Range
Error (1 year)
2V
10Hz
2.000000V
± 103.4mV
2V
1kHz
2.000000V
± 3.6mV
2V
50kHz
2.000000V
± 9.4mV
2V
100kHz
2.000000V
±14mV
2V
300kHz
2.000000V
±96mV
20V
1kHz
20.00000V
± 36mV
200V
1kHz
200.0000V
± 360mV
200mV
1kHz
200.0000mV
± 440μV
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as .
Table 3-13 Accessional AC Voltage Test ErrorDM305X
Test Signal
Input Frequency
Range
Error (1 year)
2V
10Hz
2.000000V
± 22mV
2V
1kHz
2.000000V
± 6mV
2V
50kHz
2.000000V
± 22mV
2V
100kHz
2.000000V
± 44mV
20V
1kHz
20.00000V
± 60mV
200V
1kHz
200.0000V
± 600mV
200mV
1kHz
200.0000mV
± 600μV
Notes:
[1] Set the AC filter as slow ( Meas Filter Slow );
[2] Set the reading resolution as .

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals