5.  Frequency Gain Test 
In frequency gain test, input the standard signal the test equipment outputs into 
the input terminal of tested equipment; and compare the test results with the error 
shown in the table 3-10 and 3-11. 
 
Table 3-10 Frequency Gain Error of the Standard Equipment (DM306X) 
Notes: 
[1] Use coaxial-cable as the lead for the input signal; 
[2] The reading resolution is 6½ ; 
[3] Q: Optional quick test point. 
 
Table 3-11 Frequency Gain Error of the Standard Equipment (DM305X) 
Notes: 
[1] Use coaxial-cable as the lead for the input signal; 
[2] The reading resolution is 5½ ; 
[3] Q: Optional quick test point.