RIGOL
Service Guide for DM3000 Series
3-13
5. Frequency Gain Test
In frequency gain test, input the standard signal the test equipment outputs into
the input terminal of tested equipment; and compare the test results with the error
shown in the table 3-10 and 3-11.
Table 3-10 Frequency Gain Error of the Standard Equipment (DM306X)
Test Signal
Frequency of
Test Signal
Quick Test
Point
Range
Error (1 year)
1V
5Hz
200.00000mV
3.5mHz
1V
10Hz
2.0000000V
4mHz
1V
40Hz
20.000000V
8mHz
1V
100kHz
Q
200.00000V
7Hz
Notes:
[1] Use coaxial-cable as the lead for the input signal;
[2] The reading resolution is 6½ ;
[3] Q: Optional quick test point.
Table 3-11 Frequency Gain Error of the Standard Equipment (DM305X)
Test Signal
Frequency of
Test Signal
Quick
Test Point
Range
Error (1 year)
1V
5Hz
200.00000mV
5mHz
1V
10Hz
2.0000000V
7mHz
1V
40Hz
20.000000V
8mHz
1V
100kHz
Q
200.00000V
20Hz
Notes:
[1] Use coaxial-cable as the lead for the input signal;
[2] The reading resolution is ;
[3] Q: Optional quick test point.

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals