RIGOL   
  
  Service Guide for DM3000 Series 
 
AC Voltage and AC Current Calibration 
Comparing the calibration procedure with DC, when calibrating AC function Frequency 
correction  and  Middle  range  calibration  are  appended  to  be  executed  but  Zero 
calibration. Take the calibration of AC voltage in the range of 200mV as an example:   
 
1.  Press          to select AC voltage measurement function, then press    Rng+    or 
  Rng-    to set the range as 200mV and the reading resolution as 5½ ; 
 
2.  Press          ,  and  select    T/C        PSW    to  input  password  and  set    SecrOff  ,           
then select    Cal        Enter ; 
 
3.  Select    Freq    and  select  50Hz frequency correction.  Then  input  a  sine  wave of 
100mV, 50Hz into the terminal of Input HI-LO by Fluke 5520A, after that, press 
  Meas    and select 200 kHz frequency correction. Next, input a sine wave of 200 
kHz  into  the  terminal  of  Input  HI-LO  by  Fluke  5520A.  Then  press    Meas    and 
select    Done    to return to last menu.   
 
4.  Output a sine wave of  100mV,  1 kHz into the terminal of Input HI-LO by FLuke 
5520A. Select    Middle    to perform middle range calibration, then go on output a 
sine wave of 200mV, 1kHz into the terminal of Input HI-LO by FLuke 5520A , after 
that, press    Gain    to perform full range calibration, finally select    Save    to exit 
and complete calibrating AC voltage in the range of 200mV. Afterwards, perform 
other items list in table 4-2 in the same way. 
 
 
NOTE 
Frequency correction is not required during calibrating AC current.