Service Guide for DM3000 Series
4-5
Table 4-1 DC Zero and DC Gain Calibration Input Value (DM306X)
Function
Range
Zero
DC Gain Calibration
Input Value
Resistance
200.0000Ω
0 (input terminal shorted)
200.0000Ω
2.000000kΩ
0
2.000000kΩ
20.00000kΩ
0
20.00000kΩ
200.0000kΩ
0
200.0000kΩ
10.00000MΩ
0
10.00000MΩ
100.0000
0
100.0000
DC Voltage
200.0000mV
0 (input terminal shorted)
200.0000mV
2.000000V
0
2.000000V
20.00000V
0
20.00000V
200.0000V
0
200.0000V
1000.000V
0
1000.000V
DC Current
20.00000mA
0 (input terminal open)
20.00000mA
200.0000mA
0
200.0000mA
1.000000A
0
1.000000A
10.00000A
0
10.00000A
Table 4-2 DC Zero and DC Gain Calibration Input Value (DM305X)
Function
Range
Zero
DC Gain Calibration
Input Value
Resistance
400.000Ω
0 (input terminal shorted)
400.000Ω
4.00000kΩ
0
4.00000kΩ
40.0000kΩ
0
40.0000kΩ
400.000kΩ
0
400.000kΩ
4.00000MΩ
0
4.00000MΩ
100.000MΩ
0
100.0000
DC Voltage
400.000mV
0 (input terminal shorted)
400.000mV
4.00000V
0
4.00000V
40.0000V
0
40.0000V
400.000V
0
400.000V
1000.00V
0
1000.00V
DC Current
20.0000mA
0 (input terminal open)
20.0000mA
200.000mA
0
200.000mA
1.00000A
0
1.00000A
10.0000A
0
10.0000A

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals