Service Guide for DM3000 Series
3-12
Table 3-9 AC Current Gain Error of the Standard Equipment (DM305X)
Test
Signal
Input
Frequency
Quick Test
Point
Range
Error (1 year)
20mA
1kHz
Q
20.0000mA
± 120μA
20mA
10kHz
20.0000mA
± 440μA
200mA
1kHz
Q
200.000mA
± 1.2mA
200mA
10kHz
200.000mA
± 4.4mA
1A
1kHz
Q
1.00000A
± 10mA
1A
10kHz
1.00000A
± 25mA
10A
[1]
1kHz
Q
10.0000A
± 60mA
10A
[1]
5kHz
10.0000A
± 220mA
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds ON and 30 seconds
OFF;
[2] Set the AC filter as slow ( Meas Filter Slow );
[3] Set the reading resolution as ;
[4] Q: Optional quick test point.