EasyManuals Logo

Rigol DM3000 Series Service Guide

Rigol DM3000 Series
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #53 background imageLoading...
Page #53 background image
RIGOL
Service Guide for DM3000 Series
3-11
4. AC Current Gain Test
In AC current gain test, input the standard signal the test equipment outputs into
the input terminal of tested equipment; and compare the test results with the error
shown in the table 3-8 and 3-9.
Table 3-8 AC Current Gain Error of the Standard Equipment (DM306X)
Test
Signal
Input
Frequency
Quick Test
Point
Range
Error (1 year)
20mA
1kHz
Q
20.0000mA
± 58μA
20mA
10kHz
20.0000mA
± 90μA
200mA
1kHz
Q
20.0000mA
± 580μA
200mA
10kHz
200.000mA
± 940μA
1A
1kHz
Q
1.00000A
± 5.6mA
1A
10kHz
1.00000A
± 7.5mA
10A
[1]
1kHz
Q
10.0000A
± 96mA
10A
[1]
5kHz
10.0000A
± 96mA
Notes:
[1] Continuous current more than DC 7A or AC RMS 7A should be 30 seconds ON and 30 seconds
OFF;
[2] Set the AC filter as slow ( Meas Filter Slow );
[3] Set the reading resolution as ;
[4] Q: Optional quick test point.

Other manuals for Rigol DM3000 Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DM3000 Series and is the answer not in the manual?

Rigol DM3000 Series Specifications

General IconGeneral
BrandRigol
ModelDM3000 Series
CategoryMultimeter
LanguageEnglish

Related product manuals