RIGOL
Programming Guide for DM3058
4-17
etc.), for the detailed range please refer to table 4-2.
RANGE1?: Query the range that has been selected currently on the main display.
RANGE2?: Query the range that has been selected currently on the vice display.
If the vice display does not been actived, execution error will be occur.
Return Format:
The query returns the range that has been selected currently on the main or vice
display.
Table 4-2 Optional ranges of every function
No.
DC V
ACV, Period,
Frequency
Resistance
ACI
DCI
Capacitance
0
200 mV
200 mV
20
20mA
200 μA
2nF
1
2 V
2 V
2 kΩ
200mA
2 mA
20nF
2
20 V
20 V
20 kΩ
2 A
20 mA
200nF
3
200 V
200 V
200 kΩ
10 A
200 mA
2μF
4
1000 V dc
750 V dc
2 MΩ
NA
2 A
200μF
5
NA
NA
20 MΩ
NA
10 A
10000μF
6
NA
NA
100 MΩ
NA
NA
NA
4. RATE
Command Format:
RATE<speed>
RATE?
Function:
The command sets or queries the measuring rate.
Explanations:
RATE <speed>: Set the measuring velocity as <speed> such as S (2.5
readings/s), M (20 readings/s) or F (120 readings/s). The the input state is
insensitive, any of other characters will lead to execution error.
Return Format:
The query returns the character stand for velocity: S (2.5 readings/s), M (20
readings/s) or F (120 readings/s).