etc.), for the detailed range please refer to table 4-2.
RANGE1?: Query the range that has been selected currently on the main display.
RANGE2?: Query the range that has been selected currently on the vice display.
If the vice display does not been actived, execution error will be occur.
Return Format:
The query returns the range that has been selected currently on the main or vice
display.
Table 4-2 Optional ranges of every function
4. RATE
Command Format:
RATE□<speed>
RATE?
Function:
The command sets or queries the measuring rate.
Explanations:
RATE □ <speed>: Set the measuring velocity as <speed> such as S (2.5
readings/s), M (20 readings/s) or F (120 readings/s). The the input state is
insensitive, any of other characters will lead to execution error.
Return Format:
The query returns the character stand for velocity: S (2.5 readings/s), M (20
readings/s) or F (120 readings/s).