EasyManua.ls Logo

Rigol DM3058

Rigol DM3058
264 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
RIGOL
Programming Guide for DM3058
5-26
Return Format:
The query returns max transient voltage in DCV measurement.
8. FETCh:VOLTage[:DC]:PEAK:MINimum?
Command Format:
FETCh:VOLTage[:DC]:PEAK:MINimum?
Function:
The command queries min transient voltage in DCV measurement.
Explanations:
Before sending the command, please enable DC voltage measuring function, and
enble DCV peak-peak measuring function with the command [SENSe:]
VOLTage[:DC]:PEAK:STATeON.
Return Format:
The query returns min transient voltage in DCV measurement.
9. FETCh:VOLTage[:DC]:PTPeak?
Command Format:
FETCh:VOLTage[:DC]:PTPeak?
Function:
The command queries the difference between the highest and lowest DC
transient voltage.
Explanations:
Before sending the command, please enable DC voltage measuring function, and
enable DCV peak-to-peak measuring function with the command [SENSe:]
VOLTage[:DC]:PEAK:STATeON.
Return Format:
The query returns the peak-to-peak value in DCV measurement.

Other manuals for Rigol DM3058