5-25
Before sending the command, please enable DC current measuring function, and
enble DCI peak-to-peak measuring function with the command [SENSe:]
CURRent[:DC]:PEAK:STATeON.
Return Format:
The query returns the peak-to-peak value in DCI measurement.
6. FETCh:VOLTage:AC:PTPeak?
Command Format:
FETCh:VOLTage:AC:PTPeak?
Function:
The command queries the difference between the highest and lowest AC
transient voltage.
Explanations:
Before sending the command, please enable AC voltage measuring function, and
enble ACV peak-to-peak measuring function with the command [SENSe:]
VOLTage:AC:PEAK:STATeON.
Return Format:
The query returns the peak-to-peak value in ACV measurement.
7. FETCh:VOLTage[:DC]:PEAK:MAXimum?
Command Format:
FETCh:VOLTage[:DC]:PEAK:MAXimum?
Function:
The command queries max transient voltage in DCV measurement.
Explanations:
Before sending the command, please enable DC voltage measuring function, and
enable DCV peak-to-peak measuring function with the command [SENSe:]
VOLTage[:DC]:PEAK:STATeON.