RIGOL
Programming Guide for DM3058
2-5
Function:
The command sets or queries the enable value of Status Byte Register (STB).
Explanations:
*SRE<enable_value>: Set the enable value of STB.
*SRE?: Query the enable value of STB.
<enable_value>: The range is from 0 to 188.
Return Format:
The query returns the enable value of STB.
7. *STB?
Command Format:
*STB?
Function:
The command queries the condition value of Status Byte Register (STB).
Return Format:
The query returns the condition value of STB.
8. *CLS
Command Format:
*CLS
Function:
The command clears the value from all the Event Registers and the Error Queue
information.
9. *OPC
Command Format:
*OPC
*OPC?