RIGOL
Programming Guide for DM3058
3-48
7. :trigger:ext
Command Format:
:trigger:ext?
:trigger:ext{RISE|FALL|HIGH|LOW}
Function:
The command queries and sets exterior trigger mode.
Explanations:
:trigger:ext?: Query the exterior trigger mode, and the default is the rise edge
trigger.
:trigger:ext{RISE|FALL|HIGH|LOW}:
Set the exterior trigger mode. The
parameters respectively means rise edge trigger, fall edge trigger, high level
trigger and low level trigger.
Return Format:
The query returns the current trigger mode: rise, fall, high or low.
8. :trigger:vmcomplete:polar
Command Format:
:trigger:vmcomplete:polar?
:trigger:vmcomplete:polar{POS|NEG}
Function:
The command queries and sets VMC output polar.
Explanations:
:trigger:vmcomplete:polar?: Query the VMC output polar, and the default is
positive.
:trigger:vmcomplete:polar{POS|NEG}: Set the
VMC output polar. The
parameters respectively means positive polar and negative polar.
Return Format:
The query returns the current VMC output polar: pos or neg.