Chapter 1 Specifications RIGOL
Service Guide for DM3068
1. 4 Capacitance Characteristics
Accuracy Specifications: ± (% of reading + % of range)
[1][2]
T
CAL
±5
0
to (T
CAL
-5
)
(T
+5
) to 50
Specifications are for 90 minutes warm–up and using REL operation. Additional errors may be caused by non–film capacitors.
Specifications are the 1% to 110% of range on 2nF range and 10% to 110% of range on all other ranges.
Measuring Characteristics
Apply constant current into the capacitance, and measure the voltage changing rate.
Measurement Considerations
Since small capacitance measurements are susceptible to the external noise, shielding inputs from external noise pickup is critical for minimizing
measurement errors.