RIGOL Chapter 1 Specifications
Service Guide for DM3068
1-12
1. 5 Temperature Characteristics
Accuracy Specifications
[1]
Function
Probe Type
Type
Optimum Range
1 Year
T
CAL
±5
Temperature Coefficient
0
to (T
CAL
-5
)
(T
CAL
+5) to 50
Temperature
RTD
[2]
(R
0
is within 49Ω
and 2.1kΩ)
α=0.00385
-200
to 660
0.16
0.01
α=0.00389
-200
to 660
0.17
0.01
α=0.00391
-200
to 660
0.14
0.01
α=0.00392
-200
to 660
0.15
0.01
Thermal
Resistance
2.2k
Ω
-40
to 150
0.08
0.002
3k
Ω
-40
to 150
0.08
0.002
5k
Ω
-40
to 150
0.08
0.002
10k
Ω
-40
to 150
0.08
0.002
30k
Ω
-40
to 150
0.08
0.002
Thermocouple
[3]
B
0
to 1820
0.76
0.14
E
-270
to 1000
0.5
0.02
J
-210
to 1200
0.5
0.02
K
-270
to 1372
0.5
0.03
N
-270
to 1300
0.5
0.04
R
-270
to 1768.1
0.5
0.09
S
-270
to 1768.1
0.6
0.11
T
-270
to 400
0.5
0.03
[1]
Specifications are for 90 minutes warm-up. Exclusive of sensor error.
[2]
Specification is for 4WR sensor measurement or 2WR measurement using REL operation.
[3]
Relative to cold junction temperature, accuracy is based on ITS-90. Built-in cold junction temperature refers to the temperature inside the banana jack
and its accuracy is ±2.5
.

Other manuals for Rigol DM3068