RIGOL Chapter 2 Performance Test
Service Guide for DM3068
2.5.2 Gain Test
The gain test inspects the accuracy of the meter under a full range. It is necessary
only when a regular gain offset is occurred to a function or range.
1. To test the DC gain:
(1) Make sure you have carefully read the “Test Conditions”.
(2) Connect your meter with the calibrator.
(3) Perform a step-by-step test for the functions under different ranges
specified in the following table and set the integration time to 100PLC.
Please note that none of the math function is allowable unless otherwise
stated.
(4) Input signals in accordance with the table below using the calibrator and
compare the results with the limits from the table (make sure the calibrator
can adequately output).
Table 2-4 Gain test items
Function
[1]
Range
Test
Signal
Quick
Test
[4]