RIGOL Chapter 2 Performance Test
Service Guide for DM3068
2-4
2.1.3 Routine Test
A routine test is recommended when you first obtain the meter. Please compare the
routine test results with the specifications given in the 1 year column of Allowable
Error Range from tables (Table 2-3 to Table 2-7) in Routine Test. Note the
meter should be recalibrated once the calibration interval time is up.
NOTE
The meter passed the routine test must be tested again when the test time interval
is exceeded.
The meter failed to pass the routine test must be
calibrated or repaired before it is
put back into use.