Programming Guide for DS1000/DS1000CA Series
Query Format:
:TRIGgervALTernation:TimeOFFSet?
Returned Format:
Query returns the set value of <value> in scientific numeric notation.
Example:
:TRIGger:ALTernation:TimeOFFSet□0.0002
Set up current horizontal time base as 200us
:TRIGger:ALTernation:TimeOFFSet?
Query returns 2.000e-04
6.5 :TRIGger:ALTernation<mode>:LEVel
Command Format:
:TRIGger:ALTernation<mode>:LEVel□<value>
Explanation:
The command is to set the trigger level of current channel.
<mode> can be “:EDGE”, “:PULSe” or “:VIDEO”.
The range of <value> is -6*Scale ~ + 6*Scale.
“Scale” indicates the current vertical scale. The default unit is V/div.
Query Format:
:TRIGger:ALTernation<mode>:LEVel?
Returned Format:
Query returns the set value of <value> in scientific numeric notation.
Example:
:TRIGGER:ALTERNATION:EDGE:LEVEL□2
Set the trigger level of current Channel to 2.
:TRIGGER:ALTERNATION:EDGE:LEVEL
Query returns 2.00e+00.