Chapter 2 Command System RIGOL
DS1000Z Programming Guide 2-109
:TRIGger:RUNT:BLEVel
Syntax :TRIGger:RUNT:BLEVel <level>
:TRIGger:RUNT:BLEVel?
Description Set or query the trigger level lower limit in runt trigger. The unit is the same as the current
amplitude unit.
Parameter
Name
Type
Range
Default
<level>
Real
(± 5 × VerticalScale from the screen center) - OFFSet
0
Explanation For VerticalScale, refer to :CHANnel<n>:SCALe. For OFFSet, refer
to :CHANnel<n>:OFFSet.
Return
Format
The query returns the trigger level lower limit in scientific notation.
Example
:TRIGger:RUNT:BLEVel 0.16 /*Set the trigger level lower limit to 160mV*/
:TRIGger:RUNT:BLEVel? /*The query returns 1.600000e-01*/
Related
Commands
:CHANnel<n>:SCALe
:CHANnel<n>:OFFSet