RIGOL Chapter 2 Command System
2-56 DS1000Z Programming Guide
:MEASure:ADISplay
Syntax :MEASure:ADISplay <bool>
:MEASure:ADISplay?
Description Enable or disable the all measurement function, or query the status of the all
measurement function.
Parameter
Name
Type
Range
Default
<bool>
Bool
{{0|OFF}|{1|ON}}
0|OFF
Explanation The all measurement function can measure the following 20 parameters of the
source at the same time:
Voltage Parameters: Vmax, Vmin, Vpp, Vtop, Vbase, Vamp, Vavg, Vrms, Overshoot
and Peshoot
Time Parameters: Period, Frequency, Rise Time, Fall Time, + Width, - Width, + Duty
and - Duty
Other Parameters: Area and Period Area.
The all measurement function can measure CH1, CH2, CH3 and CH4 at the same
time. You can send the
:MEASure:AMSource command to set the source of the all
measurement function.
Return
Format
The query returns 0 or 1.
Example
:MEASure:ADISplay ON
Related
Command
:MEASure:AMSource
:MEASure:AMSource
Syntax :MEASure:AMSource <src>[,<src>[,<src>[,<src>]]]
:MEASure:AMSource?
Description
Set or query the source of the all measurement function.
Parameter
Name
Type
Range
Default
<src>
Discrete
{CHANnel1|CHANnel2|CHANnel3|CHANnel4}
CHANnel1
Return
Format
The query returns CHAN1, CHAN2, CHAN3 or CHAN4.
Example
:MEASure:AMSource CHANnel1,CHANnel3