RIGOL Chapter 2 Command System
2-76 DS1000Z Programming Guide
Explanation These commands are used to select the waveform shape.
NOISe: select Noise
PULSe: select Pulse
RAMP: select Ramp
SINusoid: select Sine
SQUare: select Square
USER: select Arbitrary waveform
<freq>: set the frequency of the specified waveform (Noise does not have this
parameter). The default unit is Hz
<amp>: set the amplitude of the specified waveform. The default unit is Vpp
<offset>: set the DC offset of the specified waveform. The default unit is V
<phase>: set the start phase of the specified waveform (Noise does not have this
parameter). The default unit is degree (°).
Users can omit one or multiple parameters. When all of the parameters are omitted,
the commands only select the waveform shape but do not modify the parameters.
The four parameters (<freq>, <amp>, <offset> and <phase>) are in order. You can
not omit one parameter and set the one/ones behind it. For example, you can not
omit <freq> and set <amp> directly.

Table of Contents

Other manuals for Rigol DS1000Z Series

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1000Z Series and is the answer not in the manual?

Rigol DS1000Z Series Specifications

General IconGeneral
BrandRigol
ModelDS1000Z Series
CategoryTest Equipment
LanguageEnglish

Related product manuals