RIGOL Chapter 5 To Trigger the Oscilloscope
5-34 MSO1000Z/DS1000Z User’s Guide
I2C Trigger (Optional)
I2C is a 2-wire serial bus used to connect the microcontroller and its peripheral
device and is a bus standard widely used in the microelectronic communication
control field.
The I2C serial bus consists of SCL and SDA. Its transmission rate is determined by
SCL and its transmission data is determined by SDA, as shown in the figure below.
MSO1000Z/DS1000Z triggers on the start condition, restart, stop, missing
acknowledgement, specific device address or data value. Besides, it can also trigger
on the specific device address and data values at the same time.
Figure 5-14 Schematic Diagram of I2C Protocol
Trigger Type:
Press Type to open the trigger type list. Rotate to select "I2C" and press down
. At this point, the trigger setting information is displayed at the upper right corner
of the screen.
For example, . The trigger type is I2C trigger; the current trigger
source is CH1; the trigger level is 0.00 V.
Source Selection:
Press SCL and SDA to specify the signal sources of SCL and SDA respectively. They
can be set to CH1-CH4 or D0-D15. For the details, please refer to the introduction in
"
Trigger Source". The current trigger source is displayed at the upper right corner
of the screen.
Note: Select channel with signal input as trigger source to obtain stable trigger.
Trigger Condition:
Press When to select the desired trigger condition.
Start: trigger when SDA data transitions from high level to low level while SCL is
high level.
Restart: trigger when another start condition occurs before a stop condition.
Stop: trigger when SDA data transitions from low level to high level while SCL is
high level.
www.GlobalTestSupply.com
Find Quality Products Online at: sales@GlobalTestSupply.com

Table of Contents

Other manuals for Rigol DS1054Z

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1054Z and is the answer not in the manual?

Rigol DS1054Z Specifications

General IconGeneral
Channels4
Bandwidth50 MHz
Sample Rate1 GSa/s
Memory Depth24 Mpts
Display7 inch TFT LCD (800x480)
Vertical Sensitivity1 mV/div to 10 V/div
Time Base Range5 ns/div to 50 s/div
Vertical Resolution8 bits
Trigger TypesEdge, Pulse, Video, Slope
ConnectivityUSB Host, USB Device, LAN
Input Impedance1 MΩ
Power Supply100-240 VAC, 45-440 Hz