RIGOL Chapter 2 Command System
2-182 MSO1000Z/DS1000Z Programming Guide
:TRIGger:DURATion:TLOWer
Syntax
:TRIGger:DURATion:TLOWer <NR3>
:TRIGger:DURATion:TLOWer?
Description
Set or query the duration time lower limit in duration trigger. The default unit is s.
Parameter
Name
Type
Range
Default
<NR3>
Real
8ns to 9.99s
1μs
Explanation
This command is available when the trigger condition (:TRIGger:DURATion:WHEN) is
GREater or GLESs.
Return
Format
The query returns the duration time lower limit in scientific notation.
Example
:TRIGger:DURATion:TLOWer 0.000003 /*Set the duration time lower limit to 3μs*/
:TRIGger:DURATion:TLOWer? /*The query returns 3.000000e-06*/