RIGOL Chapter 2 Command System
2-30 MSO1000Z/DS1000Z Programming Guide
:CURSor:AUTO:AX?
Syntax
:CURSor:AUTO:AX?
Description
Query the horizontal position of cursor A in auto cursor measurement.
Explanation
 The horizontal and vertical positions of the cursor are defined by the pixel
coordinate of the screen. The pixel coordinate of the screen ranges from (0,0) to
(600,400). Wherein, (0,0) is located at the left top corner of the screen and
(600,400) is located at the right bottom corner of the screen. The horizontal pixel
range is from 0 to 600 and the vertical pixel range is from 0 to 400.
 In auto cursor measurement, cursor A of X type is not required for some of the
measurement items. At this point, the query returns 4294967295.
Return
Format
The query returns an integer between 5 and 594.
:CURSor:AUTO:BX?
Syntax
:CURSor:AUTO:BX?
Description
Query the horizontal position of cursor B in auto cursor measurement.
Explanation
 The horizontal and vertical positions of the cursor are defined by the pixel
coordinate of the screen. The pixel coordinate of the screen ranges from (0,0) to
(600,400). Wherein, (0,0) is located at the left top corner of the screen and
(600,400) is located at the right bottom corner of the screen. The horizontal pixel
range is from 0 to 600 and the vertical pixel range is from 0 to 400.
 In auto cursor measurement, cursor B of X type is not required for some of the
measurement items. At this point, the query returns 4294967295.
Return
Format
The query returns an integer between 5 and 594.
:CURSor:AUTO:AY?
Syntax
:CURSor:AUTO:AY?
Description
Query the vertical position of cursor A in auto cursor measurement.
Explanation
 The horizontal and vertical positions of the cursor are defined by the pixel
coordinate of the screen. The pixel coordinate of the screen ranges from (0,0) to
(600,400). Wherein, (0,0) is located at the left top corner of the screen and
(600,400) is located at the right bottom corner of the screen. The horizontal pixel
range is from 0 to 600 and the vertical pixel range is from 0 to 400.
 In auto cursor measurement, cursor A of Y type is not required for some of the
measurement items. At this point, the query returns 4294967295.
Return
Format
The query returns an integer between 5 and 394.

Table of Contents

Other manuals for Rigol DS1104Z-S Plus

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Rigol DS1104Z-S Plus and is the answer not in the manual?

Rigol DS1104Z-S Plus Specifications

General IconGeneral
BrandRigol
ModelDS1104Z-S Plus
CategoryTest Equipment
LanguageEnglish

Related product manuals