Chapter 2 Front Panel Operations  RIGOL 
DSG800 User Guide  2-5 
Note
[1]
: Amplitude calibration value. When calibrating the amplitude of the current 
frequency point, only the actual output amplitude will be affected and the amplitude 
display value will not change. 
 
  View the list 
You can press View to view the information of the flatness calibration list 
currently loaded. Pressing any key (except the arrow keys and knob) at the 
front panel will return to the flatness calibration list menu.   
 
Note: This menu is only valid after a list file is loaded. 
 
 
Amplitude Unit 
 
Set the unit of the RF output amplitude.   
 
Press LEVEL  Level Unit and select the desired unit from the pop-up unit menu. 
The output amplitude units available are dBm, dBmV, dBuV, Volts and Watts. Wherein, 
dBm, dBmV and dBuV are logarithmic units; Volts and Watts are linear units. The 
default is dBm.   